Search
  • Stephen Biss

34.00 +/- .05 C is manufacturer's spec


Tip 5: Manufacturer's specification for temperature stability on Guth 2100 #duisimulator is 34.00 ± .05° C., not 34.0 ± .2° C. Annual inspection, adjustment, and calibration ± .2° C. does not match manufacturer’s specs. Quality Assurance in accordance with ATC requires annual inspection, adjustment, and calibration to 34.00 ± .05° C.:

"All Approved Instruments, Approved Screening Devices and accessory equipment intended for active use in the program shall be individually inspected before being placed into service, and periodically thereafter, to ensure that they initially meet, and continue to meet the manufacturer’s specifications. The recommended interval between inspections is one year." http://www.davtech.ca/index.php/products/alcohol-testing/calibration-equipment/item/57-guth-model-2100-simulator

#manufacturersspecifications #3400 #tip

10 views0 comments